Abstract

Regular receiving coil quality assurance (QA) is required to ensure image quality of an MRIdian Linac system. The manufacturer provides a spherical phantom and positioning tube for single-slice signal-to-noise ratio (SNR) and uniformity assessments. We aimed to improve imaging setup and coverage and eliminate inter-scan variability by employing multi-slice imaging of a stable phantom. Additionally, we strived to expedite analysis by developing objective, automated analysis software. A 5300mL cylindrical plastic bottle placed in plastic bins was scanned at isocenter using a spin-echo sequence with NEMA-recommended parameters and 18 axial slices, avoiding phantom repositioning. Acquisition was repeated with and without prescan normalization filtering and by saving uncombined element images. Obtained data were analyzed using custom open-source MATLAB code. Signal and noise images were automatically assigned, and ROIs for SNR and uniformity calculations were defined using image thresholding. SNR and uniformity pass/fail decisions were made using baseline comparisons. The proposed method was successfully implemented as monthly coil QA for 3.5 years. Setup and scanning took 41 min on average for a coil set. Automated image analysis was completed in a few minutes. Signal intensity peaked around +90 or -90mm for Torso or Head/Neck coil unfiltered images. Noise peaked and minimized SNR inside±30mm from isocenter, while maximizing it around±130mm. Prescan normalization smoothed signal response, reduced SNR and increased uniformity. Individual coil element image analysis identified their position, signal or noise response and SNR. SNR and uniformity pass/fail thresholds were set for already tested and new coils. Conspicuous and subtle Torso coil malfunctions were detected considering baseline deviations of combined and individual element results. Our QA method eliminated observer bias and provided insights into coil function, image filtering performance and coil element location. It provided SNR and uniformity thresholds and identified faulty coil elements.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call