Abstract

The humidity output and CO2 elimination of the Bain circuit were tested on a simulated adult patient. The moisture content of inspired gases was found to be adequate when the circuit was used with a minute volume of 8.4 L/min and a fresh gas inflow of 4.9 L/min (65 percent relative humidity at room temperature at the onset of experimentation, rising to 100 percent after 80 minutes). However, the mean inspired CO2 concentration increased from 0.8 percent to 5.5 percent when the fresh gas inflow was decreased from 8 L/min to 3.5 L/min. It is recommended, therefore, that the circuit should not be used for long periods of time without measuring arterial CO2 tension or in situations where intentional hypocarbia is desired.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.