Abstract

In this article, it is argued that current shielding methodologies can lead to an over-shielding of low-energy X-ray imaging facilities. Measurements of scattered air kerma rates in a mammography installation, including the effect of an anthropomorphic phantom of the patient, are presented. Detailed Monte Carlo calculations that support the measured results are also shown. Calculations for a mobile mammography installation, including the 'patient transmission factor', are presented and results are discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.