Abstract

This study reviews the literature regarding the factors contributing to failures of dental implants. It may be suggested that the following situations are correlated to increase the implant failure rate: a low insertion torque of implants that are planned to be immediately or early loaded, inexperienced surgeons inserting the implants, implant insertion in the maxilla, implant insertion in the posterior region of the jaws, implants in heavy smokers, implant insertion in bone qualities type III and IV, implant insertion in places with small bone volumes, use of shorter length implants, greater number of implants placed per patient, lack of initial implant stability, use of cylindrical (nonthreaded) implants and prosthetic rehabilitation with implant- supported overdentures. Moreover, it may be suggested that the following situations may be correlated with an increase in the implant failure rate: use of the non- submerged technique, immediate loading, implant insertion in fresh extraction sockets, smaller diameter implants. Some recently published studies suggest that modern, moderately rough implants may present with similar results irrespective if placed in maxillas, in smoking patients or using only short implants.