We have developed a clinically compatible, real-time ultrasound needle tracking system (UNT) that can be appended to a clinical ultrasound system, superimposing a crosshair onto the ultrasound image at the needle tip position. The UNT was developed under the ISO 13485 Medical Devices quality standard for deployment in the clinic. During handheld ultrasound guidance, the location of the needle tip within the imaging plane is determined from the acoustic signals received by an embedded fibre-topic hydrophone. Assessment of tracking accuracy found that the mean distance between tracked and true positions was 0.7 ± 0.4 mm with a repeatability of 0.3 ± 0.2 mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.