Abstract

Advanced high quality radiation therapy techniques such as IMRT require an accurate delivery of precisely modulated radiation fields to the target volume. Interfractional and intrafractional motion of the patient's anatomy, however, may considerably deteriorate the accuracy of the delivered dose to the planned dose distributions. In order to compensate for these potential errors, a dynamic real-time capable MLC control system was designed. The newly developed adaptive MLC control system contains specialized algorithms which are capable of continuous optimization and correction of the aperture of the MLC according to the motion of the target volume during the dose delivery. The algorithms calculate the new leaf positions based on target information provided online to the system. The algorithms were implemented in a dynamic target tracking control system designed for a Siemens 160 MLC. To assess the quality of the new target tracking system in terms of dosimetric accuracy, experiments with various types of motion patterns using different phantom setups were performed. The phantoms were equipped with radiochromic films placed between solid water slabs. Dosimetric results of exemplary deliveries to moving targets with and without dynamic MLC tracking applied were compared in terms of the gamma criterion to the reference dose delivered to a static phantom. Our measurements indicated that dose errors for clinically relevant two-dimensional target motion can be compensated by the new control system during the dose delivery of open fields. For a clinical IMRT dose distribution, the gamma success rate was increased from 19% to 77% using the new tracking system. Similar improvements were achieved for the delivery of a complete IMRT treatment fraction to a moving lung phantom. However, dosimetric accuracy was limited by the system's latency of 400 ms and the finite leaf width of 5 mm in the isocenter plane. Different experimental setups representing different target tracking scenarios proved that the tracking concept, the new algorithms and the dynamic control system make it possible to effectively compensate for dose errors due to target motion in real-time. These early results indicate that the method is suited to increasing the accuracy and the quality of the treatment delivery for the irradiation of moving tumors.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.