Abstract
Left atrial radiofrequency catheter ablation (RFA) is gaining acceptance as treatment for drug-refractory atrial fibrillation (AF). This therapy has been associated with esophageal injury and atrioesophageal fistula formation causing death. We describe 3 patients undergoing catheter ablation for AF during real-time monitoring of luminal esophageal temperature. We observed heating of the esophagus during short duration low power RFA, at either the left or right pulmonary vein ostia. Cryoablation at the pulmonary vein ostium in one patient resulted in esophageal cooling. Furthermore, we observed that fluoroscopic localization of the ablation catheter at a site apparently distant from the esophagus is not adequate to assure avoidance of ablation-induced esophageal heating. Real-time monitoring of luminal esophageal position and temperature is feasible, enhances recognition of esophageal heating, and may add useful information beyond that provided by fluoroscopic assessment of esophageal position. There is a potential role for esophageal monitoring to help avoid thermal injury to the esophagus during catheter ablation for atrial fibrillation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.