Abstract

Nucleic acid amplification reactions such as polymerase chain reaction (PCR), which uses a DNA polymerase to amplify individual double-stranded DNA fragments, are a useful technique for visualizing the presence of specific genomes. Although the fluorescent labeling method is mainly used with DNA amplification, other detection methods should be considered for further improvements, such as miniaturization and cost reduction, of reaction-monitoring devices. In this study, the quartz-crystal microbalance (QCM) method, which can measure nanogram-order masses, was applied for the real-time detection of DNA fragments in a solution with nucleic acids. This was combined with an isothermal nucleic acid amplification reaction based on the recombinase polymerase amplification (RPA) method, which allowed DNA amplification at a constant temperature. When the DNA amplification reaction was initiated on a QCM sensor plate with an immobilized primer DNA strand, a significant increase in mass was observed compared to when the primer DNA was not immobilized. QCM was shown to be sufficiently sensitive for the in situ detection of amplified DNA fragments. Combining a portable QCM device and RPA offers a sensitive point-of-care method for detecting nucleic acids.