Abstract

Near-infrared photoimmuno-therapy (NIR-PIT) is a newly approved cancer phototherapy. We aim to confirm whether a clinically approved camera for indocyanine green (ICG) could monitor IR700 fluorescence in real time during exposure to therapeutic NIR light. An NIR camera, LIGHTVISION, designed to image ICG fluorescence, was used. A431-GFP/luc tumor-bearing mice were exposed to therapeutic NIR light and real-time fluorescence imaging (RT-FI) was obtained and measured with LIGHTVISION. Bioluminescence imaging (BLI) was performed to confirm cell death. RT-FI during NIR-PIT revealed an initial rapid loss of fluorescence, followed by a plateau which occurred at a light dose of approximately 30 J/cm 2 . Correlation between BLI and IR700 fluorescence loss showed that loss of fluorescence was associated with increased cell death. The efficacy of NIR-PIT could be monitored non-invasively and in real-time using weak fluorescence at wavelengths much longer than the peak fluorescence of IR700. This technique can achieve precise light dosimetry that allows us to decide on the optimal exposure.