Abstract
The objective of this work is to document performance of automated insulin delivery (AID) during real-life use in type 2 diabetes (T2D). A retrospective analysis was performed of continuous glucose monitoring and insulin delivery data from 796 individuals with T2D, who transitioned from 1-month predictive low-glucose suspend (PLGS) use to 3-month AID use, in real-life settings. Primary outcome was change of time in range (TIR = 70-180 mg/dL) from PLGS to AID. Secondary outcomes included time above/below range (TAR/TBR) and total daily insulin (TDI). Compared with PLGS, AID increased TIR on average from 63.2% to 72.6%, decreased TAR from 36.2% to 26.8%, and increased TDI from 70.2 to 76.3 U (all P < .001), without significant change to TBR. Glycemic improvements were more pronounced in those with worse glycemic control during PLGS use (P < .001). Real-life use of AID led to a rapid and sustained improvement of glycemic control in individuals with T2D.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.