Abstract
We present the concept of and the procedure for realizing the primary scale of the spectral responsivity based on an absolute cryogenic radiometer at the Korea Research Institute of Standards and Science (KRISS). A detailed error and uncertainty analysis is performed by calibrating the standard detectors in the wavelength range from 300 nm to 1000 nm. The relative expanded uncertainty of the spectral responsivity calibration for a Si trap detector is 0.06% at 632.8 nm and 0.4% for other wavelengths (k = 2). The calibration results are validated by comparing with the results from a theoretical model of quantum efficiency for a transmission-type trap detector.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Similar Papers
More From: Journal of the Korean Physical Society
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.