Abstract

This article summarizes data from a 3-year, double-blinded study of directional hearing aid benefit. Ninety-four subjects in three hearing loss groups, all previous users of omnidirectional output-compression hearing aids, completed all aspects of the study. Participants were fit with new hearing aids for 1 month in a directional mode and 1 month in an omnidirectional mode. Following 1 month of use, subjects completed a number of objective and subjective measures of hearing aid outcome. Objective and subjective data were analyzed across hearing aid and hearing loss conditions. Subjects in all hearing loss groups exhibited better performance in the directional conditions for objective speech-in-noise measures; however, subjective data did not indicate a clear advantage for directional amplification. Results and clinical implications are discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.