Abstract

Noninvasive thermal therapies for the treatment of breast cancer depend on accurate monitoring of tissue temperature to optimize treatment and ensure safety. This work describes a real-time system for 3-D thermoacoustic imaging and thermometry (TAI-TAT) for tracking temperature in tissue samples during heating. The study combines a 2.7-GHz microwave pulse generator with a custom 1.5-D 0.6 MHz ultrasound array for generating and detecting TA signals. The system is tested and validated on slabs of biological tissue and saline gel during heating. Calibration curves for relating the TA signal to temperature were calculated in saline gel (3.40%/°C), muscle (1.73%/°C), and fat (1.15%/°C), respectively. The calibrations were used to produce real-time, volumetric temperature maps at ~3-s intervals with a spatial resolution of approximately 3 mm. TAT temperature changes within a region of interest were compared to adjacent thermocouples with a mean error of 17.3%, 13.2%, and 20.4% for muscle, gel, and fat, respectively. The TAT algorithm was also able to simultaneously track temperatures in different tissues. With further development, noninvasive TAI-TAT may prove to be a valuable method for accurate and real-time feedback during breast cancer ablation therapy.