Abstract

Using catheter-mounted 2-D array transducers, we have obtained real-time 3-D intracardiac ultrasound (US) images. We have constructed several transducers with 64 channels inside a 12 French catheter lumen operating at 5 MHz. The transducer configuration may be side-scanning or beveled, with respect to the long axis of the catheter lumen. We have also included six electrodes to aquire simultaneous electrocardiograms. Using an open-chest sheep model, we inserted the catheter into the cardiac chambers to study the utility of in vivo intracardiac 3-D scanning. Images obtained include a cardiac four-chamber view, mitral valve, pulmonic valve, tricuspid valve, interatrial septum, interventricular septum and ventricular volumes. We have also imaged two electrophysiological interventional devices in the right atrium, performed an in vitro ablation study, and viewed the pulmonary veins in vitro. (E-mail: edl@duke.edu)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.