Dehydration is one of the most profound physiological challenges that significantly affects athletes and soldiers if not detected early. Recently, a few groups have focused on dehydration detection using sweat as the main biomarker. Although there are some proposed devices, the electrical and chemical characteristics of sweat have yet to be incorporated into the validations. In this work, we have developed a simple test setup to analyze artificial sweat that is comprised the main components of human sweat. We provide theoretical and experimental details on the electrical and chemical behavior of the artificial sweat for various concentration values within a temperature range of 5 °C to 50 °C. We have also developed an efficient sweat collecting and detection system based on 3D printing. Human studies were conducted and this particular protocol has shown that dehydration starts to take effect as early as 40 min into the physical activity if there is no fluid intake during the exercise. We believe that our device will lead to developing viable real-time sweat analysis systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.