Abstract
Current real-time volumetric scanners use a 2-D array to scan a pyramidal volume consisting of many sector scans stacked in the elevation direction. This scan format is primarily useful for cardiac imaging to avoid interference from the ribs. However, a real-time rectilinear volumetric scan with a wider field of view close to the transducer could prove more useful for abdominal, breast, or vascular imaging. In previous work, computer simulations of very sparse array transducer designs in a rectilinear volumetric scanner demonstrated that a Mills cross array showed the best overall performance given current system constraints. Consequently, a 94 x 94 Mills cross array including 372 active channels operating at 5 MHz has been developed on a flexible circuit interconnect. In addition, the beam former delay software and scan converter display software of the Duke volumetric scanner were modified to achieve real-time rectilinear volumetric scanning consisting of a 30-mm x 8-mm x 60-mm scan at a rate of 47 volumes/s. Real-time rectilinear volumetric images were obtained of tissue-mimicking phantoms, showing a spatial resolution of 1 to 2 mm. Images of carotid arteries in normal subjects demonstrated tissue penetration to 6 cm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.