Abstract
In this paper we report on Monte Carlo simulations to investigate real-time monitoring of the track depth profile in particle therapy by measuring prompt gamma ray emissions: a high sensitivity imaging system employing a knife-edge-shaped slit combined with a position-sensitive gamma detector was evaluated. Calculations to test this new concept were performed for a head-sized software phantom. Clear spatial correlation is shown between the distribution of gamma rays detected with energies above 1.5 MeV and the distribution of prompt gamma rays emitted from the phantom. The number of neutrons originating from nuclear reactions in the phantom that are detected at these high energies is small. Most importantly it is shown that under common therapy conditions enough data may be collected during one spot-step (of the order of 10 ms) to locate the distal dose edge with a 1σ accuracy of better than 1 mm. This indicates that simple slit cameras have high potential for accurate real-time particle therapy adjustment and may become a practical way to improve particle therapy accuracy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.