Abstract

We developed a real-time multi-channel Fourier transform spectrometer with an area sensor to track biological sample motion. In order to decrease the time intervals between measurements, a simultaneous measurement system consisting of two interferograms with opposite phases was constructed by using an area sensor instead of the often-used linear sensor. Since this system requires no mechanical motion for measurement, the rate at which it outputs spectra depends only upon the frame rate of the area sensor. This spectrometer was employed in the near-infrared region at wavelengths of 900–1700nm to realize non-invasive blood testing. Substances in blood can be measured at 20Hz using this method, which is higher than the pulse rate of a human blood vessel. This high-frequency technique enables the extraction of only the blood spectra from samples obtained from fingertips. In this study, the concentration of neutral fat within the blood was determined by extracting the periodically changing signal from the 1200nm peak.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.