Abstract

Purpose To develop interventional magnetic resonance (MR) guidance techniques for inferior vena cava (IVC) filter retrieval in vitro and demonstrate feasibility in vivo. Materials and Methods Three optional IVC filters and their retrieval systems were investigated. Experiments were performed on a 1.5-T MR system. Real-time MR imaging was optimized by using a custom-built IVC phantom. A three-dimensional (3D) contrast-enhanced MR venography sequence was optimized in vitro for improved detection of thrombus trapped within the filters. Filters were then retrieved in vitro and in vivo in a swine model under MR guidance. In-vivo retrieval procedure time was measured. Results The combination of one of the nitinol filters and a loop snare was suitable for real-time MR procedures. With a 90° flip angle, 3D MR venography allowed detection of simulated thrombus within the filter. A radial true fast imaging sequence with steady-state precession allowed visualization of the loop snare and IVC filter hook and successful retrieval of the filter in vivo and in vitro. In-vivo MR fluoroscopy time for retrieval was 97 seconds ± 51 (mean ± SD). Conclusions MR-guided retrieval of a nitinol-based IVC filter by using a loop snare is feasible with the use of optimized sequences and passive device tracking.