A new intravascular fluorosensor was developed and validated for inline P(CO)₂monitoring. The P(CO)₂sensor was based on the fluorescent indicator 1-hydroxypyrene-3,6,8-trisulfonate. The P(CO)₂sensor was then immersed in various solutions in vitro and carotid artery bypass of rabbits in vivo for testing. Changes of P(CO)₂in solutions and blood were created by bubbling CO2/N2 and hyperventilation/hypoventilation, respectively. The changes of fluorescent intensity over P(CO)₂ range of 14-150 mmHg was linear. The resolution of the whole sensor system was 1 mmHg, with a bias ± SD of -0.1 ± 2.9 mmHg and precision ± SD of 2.1 ± 1.9 mmHg. The sensor signal has been stable during measurement for at least 25h and was insensitive to fluctuations of ions concentration and osmosis at pathophysiological limits. The performance of the sensor is in agreement with blood gas analyzer in a wide range of P(CO)₂and it is qualified for continuous intravascular measurement of blood P(CO)₂at various conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.