Abstract

A method is developed for measuring the electroosmotic (bulk) flow rate continuously in capillary zone elctrophoretic separations, which is based on introducing a reference flow of solution containing a fluorescent marker into the running buffer downstream from the detection zone. This technique can determine changes as small as 1% in the flow rat during a separation. The response time of the measurement is ∼1s, and the dynamic range extends over more than an order of magnitude in flow rate

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.