Abstract
There is an increasing interest in new strategies to detect neurotransmitters released from nerve cells in real time for brain science, drug assessment, and so on. Previously we reported real-time monitoring of dopamine release from nerve model cells by enzyme-catalyzed luminescence measurement with tyramine oxidase and peroxidase. In the present study, the system was modified with glutamate oxidase instead of tyramine oxidase to detect L-glutamate sensitively ( approximately 10 nM) and rapidly with high temporal resolution (<1 s). We applied this modified method successfully to perform real-time monitoring of L-glutamate release from brain model cell (C6 glioma cell) using a luminescence plate reader upon stimulation with high concentration of KCl (>10 mM) or 5-hydroxytryptamine (>1 microM). The measurement solution was not toxic and therefore the L-glutamate release from the cell was measured by the second stimulation after exchanging the measurement solution. We conclude that the developed monitoring system is suitable for real-time detection of dynamic L-glutamate release from nerve cells in vitro and will be suitable for application in assessment of drugs acting on the nervous system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.