Abstract

We designed a ventilation-perfusion index (VQI) to estimate venous admixture (Qsp/Qt) in a real-time fashion by simultaneous pulse and pulmonary artery oximetry in 17 patients with acute respiratory failure. Changes in Qsp/Qt were produced by altering the level of continuous positive airway pressure. VQI correlated well with Qsp/Qt (r = 0.78). This contrasts with the poor correlation found between Qsp/Qt and the commonly used oxygen tension based indices such as PaO2/FIO2 (r = -0.51), PaO2/PAO2 (r = 0.47), and PAO2-PaO2 (r = 0.23). The use of dual oximetry to derive a VQI appears to be a reliable and accurate method for real-time assessment of pulmonary gas exchange in patients with acute respiratory failure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.