Abstract

Sufficient energy resolution is the key issue for the calorimetry in particle and nuclear physics. The calorimeter of the A4 parity violation experiment at MAMI is a segmented calorimeter where the energy of an event is determined by summing the signals of neighboring channels. In this case, the precise matching of the individual modules is crucial to obtain a good energy resolution. We have developed a calibration procedure for our total absorbing electromagnetic calorimeter which consists of 1022 lead fluoride (PbF 2) crystals. This procedure reconstructs the single-module contributions to the events by solving a linear system of equations, involving the inversion of a 1022×1022-matrix. The system has shown its functionality at beam energies between 300 and 1500 MeV and represents a new and fast method to keep the calorimeter permanently in a well-calibrated state.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.