Abstract
Thermoacoustic (TA) imaging is a modality where pulsed microwaves are used to generate ultrasound waves in tissue, which are highly correlated with temperature. This study uses a self-calibration approach to improve the estimation of temperature using 3D real-time TA thermometry in porcine tissue during localized heating. The self-calibration method estimated temperatures at eight embedded thermocouple locations with a normalized root-mean-square error of 3.25±2.08%. The results demonstrate that the method has the suitable accuracy and resolution to provide feedback control for breast cancer ablation therapy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.