Abstract

Thermoacoustic (TA) imaging is a modality where pulsed microwaves are used to generate ultrasound waves in tissue, which are highly correlated with temperature. This study uses a self-calibration approach to improve the estimation of temperature using 3D real-time TA thermometry in porcine tissue during localized heating. The self-calibration method estimated temperatures at eight embedded thermocouple locations with a normalized root-mean-square error of 3.25±2.08%. The results demonstrate that the method has the suitable accuracy and resolution to provide feedback control for breast cancer ablation therapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.