Abstract

Combination of electronic portal imaging device (EPID) transmission imaging with frontal Cherenkov imaging enabled real-time 3D dosimetry of clinical X-ray beams in water phantoms. The EPID provides a 2D transverse distribution of attenuation which can be back-projected to estimate accumulated dose, while the Cherenkov image provides an accurate lateral view of the dose versus depth. Assuming homogeneous density and composition of the phantom, both images can be linearly combined into a true 3D distribution of the deposited dose. We describe the algorithm for volumetric dose reconstruction, and demonstrate the results of a volumetric modulated arc therapy (VMAT) 3D dosimetry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.