Abstract

IntroductionLaboratory test interferences can cause spurious test results and patient harm. Knowing the frequency of various interfering substances in patient populations likely to be tested with a particular laboratory assay may inform test development, test utilization and strategies to mitigate interference risk. MethodsWe developed REACTIR (Real Evidence to Assess Clinical Testing Interference Risk), an approach using real world data to assess the prevalence of various interfering substances in patients tested with a particular type of assay. REACTIR uses administrative real world data to identify and subgroup patient cohorts tested with a particular laboratory test and evaluate interference risk. ResultsWe demonstrate the application REACTIR to point of care (POC) blood glucose testing. We found that exposure to several substances with the potential to interfere in POC blood glucose tests, including N-acetyl cysteine (NAC) and high dose vitamin C was uncommon in most patients undergoing POC glucose tests with several key exceptions, such as burn patients receiving high dose IV-vitamin C or acetaminophen overdose patients receiving NAC. ConclusionsFindings from REACTIR may support risk mitigation strategies including targeted clinician education and clinical decision support. Likewise, adaptations of REACTIR to premarket assay development may inform optimal assay design and assessment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.