Abstract

In order to determination of cholesterol in human serum with no reagent using near-infrared (NIR) spectroscopy. Interval partial least square (iPLS) was proposed as an effective variable selection approach for multivariate calibration. For this purpose, an independent sample set was employed to evaluate the prediction ability of the resulting model. The spectrum was split into different interval. Then, the informative region of cholesterol (1688-1760nm), in which the PLS model has a low RMSEP with 0.241mmol/L and a high R with 0.975, is selected with 23 intervals. The results indicate that, the informative region of cholesterol can be obtained by iPLS and applied to design the simpler reagentless NIR instruments for inexpensive cholesterol measurement in future.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.