Abstract

The readout electronics for a prototype soft X-ray spectrometer based on silicon drift detector (SDD), for precisely measuring the energy and arrival time of X-ray photons is presented in this paper. The system mainly consists of two parts, i.e., an analog electronics section (including a pre-amplifier, a signal shaper and filter, a constant fraction timing circuit, and a peak hold circuit) and a digital electronics section (including an ADC and a TDC). Test results with X-ray sources show that an energy dynamic range of 1–10 keV with an integral nonlinearity of less than 0.1% can be achieved, and the energy resolution is better than 160 eV @ 5.9 keV FWHM. Using a waveform generator, test results also indicate that time resolution of the electronics system is about 3.7 ns, which is much less than the transit time spread of SDD (<100 ns) and satisfies the requirements of future applications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.