Abstract
We have shown that several commercial control sera containing reversibly inactivated creatine kinase are reactivated by pre-incubation in N-acetyl cysteine assay diluent. The increase in activity can proceed for up to 4 h. This reactivation is not seen in patients' sera even after storage for 14 days at -20 degrees C, 7 days at 4 degrees C, or 2 days at 20 degrees C, during which time no loss of activity was evident. This study brings into question the validity of using consensus values to standardise enzyme assays.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annals of Clinical Biochemistry: International Journal of Laboratory Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.