Abstract

The aim of this study was to evaluate image quality and radiation dose in low-dose head and neck CT comparing two different commercially available iterative reconstruction algorithms: sinogram-affirmed iterative reconstruction (SAFIRE) and advanced modeled iterative reconstruction (ADMIRE) with fixed and automated tube voltage adaptation (TVA). CT examinations of 103 patients were analysed. 58 patients were examined on a single-source CT at fixed tube voltage of 120 kV and reconstructed with filtered back projection (FBP) and SAFIRE (Strength Level 3). 45 patients were examined in a single-source mode on a dual-source CT with automated TVA and reconstructed with FBP and ADMIRE (Strength Levels 2 and 3). Image noise was calculated in seven anatomical volumes of interest. Subjective evaluation of the CT images was performed using a four-grade scale. Mean CT numbers of FBP and the corresponding iterative reconstruction did not differ significantly (p = 0.74-0.99). Image noise was lower with both iterative reconstruction techniques than with FBP (SAFIRE 3: -22.3%; ADMIRE 2: -14.9%; ADMIRE 3: -24.2%; all p < 0.05); hence, the signal-to-noise ratio and the contrast-to-noise values were higher. Subjective image quality revealed a more favourable result for the iterative reconstruction. ADMIRE 3 in combination with automated TVA showed 14.4% (p < 0.05) less image noise with a 7.5% lower radiation dose than SAFIRE 3 with fixed tube voltage. Higher image quality at lower radiation dose can be achieved using ADMIRE in combination with automated TVA.