Abstract

ABSTRACT Purpose Longitudinally verify the influence of auditory tonal thresholds obtained with transcutaneous and percutaneous bone-anchored hearing aids on speech perception in individuals with external and/or middle ear malformation and chronic otitis media. Methods Observational, retrospective, longitudinal follow-up study of 30 unilateral users of the transcutaneous and percutaneous Baha® system for the collection of secondary data on pure tone thresholds obtained through free field audiometry and sentence recognition threshold in silence and noise in conditions: without the prosthesis; at the time of activation; in the first month of use (post 1); and in the third month (post 2). Results There was a significant difference between pure tone thresholds obtained at frequencies of 3 and 4kHz with better results for the percutaneous technique at all evaluation moments. For both systems, better performance was observed in sentence recognition in silence and in noise, with a significant difference in activation (p<0.001), but it remained stable during the other evaluation moments. The percutaneous system showed better benefit in recognizing sentences in noise only on activation (p=0.036), when compared to the transcutaneous system. Conclusion The percutaneous system provided better audibility for high frequencies; however, such audibility did not influence sentence recognition in the silent situation for both systems. For the noise situation, better responses were observed in the percutaneous system, however, the difference was not maintained over time.