Abstract

Aim of Study: With an ab-interno technique of transscleral suturing of current one-piece posterior chamber intraocular lenses (PC IOLs) by injector implantation in the absence of capsular support, we aimed to demonstrate the possibility of the implantation of one-piece acrylic PC IOLs that might be produced in the future for only scleral fixation through small clear corneal incision. Materials and Methods: Case report and literature review. Results: This procedure has been performed in eight aphakic eyes with four different types of IOLs. Good centration was achieved with minimal technical effort. All patients had well-centered and stable lenses postoperatively during 9-18 months follow-up. Conclusion: We managed to decrease the risks of surgical trauma and intricate surgical maneuvers requirement. With this technique, excessive fluid leakage and consecutive hypotony can be minimized.