Abstract

Based on the single chip microcomputer (SCM) control theory, the control circuit of magnetic driving system of axial flow blood pump is designed. For the easy damage of TIP 122, which is used as the power switch of control circuit, RC snubber circuit is designed by analysis and experiments. And then an optimal matching of the key parameters is obtained to solve the security problem of the TIP 122. The experimental results show that the RC snubber circuit can improve the reliability and stability of the driving system, and to provide an experimental foundation for the further studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.