Abstract
Fracture of nickel-titanium rotary files is an iatrogenic error which can seriously jeopardize root canal therapy. If a high-torque motor is used, the instrument-specific limit-torque (fracture limit) is often exceeded, thus increasing the risk of intracanal failure. A possible solution to this problem is to use a low-torque endodontic motor which operates below these values. If the torque is set just below the limit of elasticity for each instrument, the risk of fracture is likely to be markedly reduced. The purpose of this paper was to discuss mechanical properties of NiTi rotary instruments, the rationale for selecting low torque values, and to use clinically a new endodontic motor (step-motor) which operates below the limit of elasticity of each rotary file. The step-motor was found to be helpful in reducing the risk of instrument fracture. Irreversible material damage (plastic deformation) and instrument fracture were rarely seen. Low-torque instrumentation also increased tactile sense and, consequently, mental awareness of rotary instrumentation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.