Abstract
Chronotropic integrity is required for a normal cardiac output response to exercise. We evaluated a rate-adaptive ventricular demand pacemaker (Telectronics, META-MV) which uses minute ventilation as the sensed physiological variable for adjusting pacing rate, in seven young patients with a mean age of 11.4 years. All patients had clinically significant bradycardia related to complete heart block (n = 4) or sinus node dysfunction (n = 3). For the entire group, paced heart rates increased from 70 +/- 10 beats/min to 151 +/- 19 beats/min with exercise testing. The onset of rate adaptation took less than 30 seconds. Changes in paced rate were linearly related to workload, VO2 (5.9 to 20.7 mL/min/kg) and minute ventilation (8-65 L/min). The decline in pacing rate after exercise was related directly to the gradual decrease in minute ventilation and VO2. Our data show that minute ventilation closely and accurately reflects the metabolic demands of varying workloads in children and can be used to achieve physiological, rate-adaptive pacing.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.