Abstract
Most fast-imaging sequences use gradient echoes and a low flip-angle excitation. The low flip angle is used because it gives increased signal when TR < T 1. However, spin-echo sequences are less productive of certain artifacts, among them flow and magnetic susceptibility artifacts. We present a modification of the spin-echo pulse sequence designed to optimize the signal-to-noise ratio for repetition times (TR) less than 100 msec while preserving good image quality. Our implementation performs a 128 × 128 image in under 7 sec (TR = 50 msec) and has been used to follow the dynamics of Gd-DTPA in the rat kidneys.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.