Abstract

ObjectivesThe ongoing global severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) pandemic necessitates adaptations in the practice of surgical pathology at scale. Primary diagnosis by whole-slide imaging (WSI) is a key component that would aid departments in providing uninterrupted histopathology diagnosis and maintaining revenue streams from disruption. We sought to perform rapid validation of the use of WSI in primary diagnosis meeting recommendations of the College of American Pathologists guidelines.MethodsGlass slides from clinically reported cases from 5 participating pathologists with a preset washout period were digitally scanned and reviewed in settings identical to typical reporting. Cases were classified as concordant or with minor or major disagreement with the original diagnosis. Randomized subsampling was performed, and mean concordance rates were calculated.ResultsIn total, 171 cases were included and distributed equally among participants. For the group as a whole, the mean concordance rate in sampled cases (n = 90) was 83.6% counting all discrepancies and 94.6% counting only major disagreements. The mean pathologist concordance rate in sampled cases (n = 18) ranged from 90.49% to 97%.ConclusionsWe describe a novel double-blinded method for rapid validation of WSI for primary diagnosis. Our findings highlight the occurrence of a range of diagnostic reproducibility when deploying digital methods.