Abstract

A rapid tandem mass spectrometric (MS-MS) method for the quantification of gabapentin (GBP) in human plasma using 4-phenyl-4-aminobutanoic acid as an internal standard (IS) has been developed and validated. The drug and the internal standard were analyzed, by flow injection analysis without chromatographic separation, using a mobile phase of acetonitrile-water-formic acid (50:50:0.025, v/v/v) at a flow rate of 0.1 mL min−1. The run-cycle time was 154 and m/z 180 > 117 for GBP and the IS, respectively. Ion suppression study indicated practically no suppressive effect of plasma constituents on the mass ions detection of GBP and IS, when measured in MRM scanning mode. Calibration curves were linear over the concentration range of 0.1–10 μg mL−1 (r > 0.999) with a limit of quantification of 0.1 μg mL−1 (RSD%; 7.6 and % DEVs; −3.0 to +17.0%). Validation data showed that the RSD% values were in the range of 1.85 to 13.06%, whereas, the % DEVs values ranged from −1.4 to +10.0% indicating good precision and accuracy. Analytical recoveries of GBP from spiked human plasma were in the range of 98.9 to 101.3%. On the other hand, recoveries of GBP from stored human plasma samples were in the range of 100.0 to 107.5% indicating that GBP was stable in plasma, with no appreciable degradation, when stored at −20 °C. The developed method was applied for GBP monitoring in plasma samples of patients treated with GBP.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.