Abstract

A rapid screening assay (9 min/sample) has been developed and validated for the detection of deoxynivalenol in durum wheat, wheat products, and maize-based baby foods using an SPR biosensor. Through a single laboratory validation, the limits of detection (LOD) for wheat, wheat-based breakfast cereal, and maize-based baby food were 57, 9, and 6 μg/kg, respectively. Intra-assay and interassay precisions were calculated for each matrix at the maximum and half-maximum European Union regulatory limits and expressed as the coefficient of variation (CV). All CVs fell below 10% with the exception of the between-run CV for breakfast cereal. Recoveries at the concentrations tested ranged from 92 to 115% for all matrices. Action limits of 161, 348, and 1378 μg/kg were calculated for baby food, wheat-based breakfast cereal, and wheat, respectively, and the linear range of the assay was determined as 250-2000 μg/kg.