Abstract
Abstract We established a highly reproducible, simple and rapid simultaneous measurement method in which blood plasma can be directly injected into high-performance liquid chromatography (HPLC) without conducting conventional extraction of kynurenine (Kyn) and tryptophan (Trp) from blood plasma. In this method, Kyn and Trp were first separated from proteins present in blood plasma using a pre-column. Then, by column-switching, the separated Kyn and Trp were automatically injected into a separation column. The recoveries of Kyn and Trp determined in blood plasma by this method were 100.0 ± 0.9 and 100.0 ± 1.4%, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.