Abstract
A synchronised flow system with hydride generation coupled to ICP-MS is proposed for the sequential determination of As and Se in natural waters and plant digests. The alternated mixing of the sample solution with thiourea or HCl for the determination of As or Se under optimized conditions was achieved using a flow commutator before the reaction with NaBH 4. The on-line addition of thiourea promoted the quantitative reduction of As(V) to As(III), thus enhancing sensitivity and precision. The selenium pre-reduction from Se(VI) to Se(IV) was produced by heating the sample with HCl, and the hydride generation was performed in 4 mol l −1 HCl, thus avoiding interference from thiourea. The system allowed the analysis of 20 samples h −1 with LOD values of 0.02 μg l −1 As and 0.03 μg l −1 Se. Results were in agreement with the certified values at the 95% confidence level for reference waters from the Canadian National Water Research Institute and plant samples from the National Institute of Standards and Technology (NIST).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.