Abstract
An electrospray ionization (ESI) ambient pressure ion-mobility spectrometer (APIMS) interfaced to an orthogonal reflector time-of-flight mass spectrometer (TOFMS) was evaluated for the first time as a detector for the identification of phenylthiohydantoin (PTH)-derivatized amino acids, the final products in the Edman sequencing process of peptides and proteins. The drift and flight times of the twenty common PTH amino acids were characterized by a well-defined 2-D mobility/mass spectral pattern. The combination of mobility/mass modes of analysis gave rise to a unique trend-line formation for the series of PTH amino acids. In addition, each PTH amino acid had a unique reduced mobility constant K(o), thus enabling the differentiation of all the amino acid derivatives including the PTH-leucine and PTH-isoleucine isomers. More importantly it was shown that it was possible to resolve a complete reference mixture of PTH amino acids in a single experimental run in less than 1 min. Detection limits for the PTH amino acids were found to range from 1.04 to 3.52 ng; indicating that the limits of detection were less than 17.0 pmol for all of the PTH amino acids.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.