Abstract

Simple and rapid reverse phase high-performance liquid chromatography (RP-HPLC) method was developed and validated using Phenomenex Gemini c18 (4.6 x 250 mm, 5 µ) reverse phase column for the determination of LAF in human plasma, Solid Phase Extraction (SPE) technique was used for the extraction of analyte, detection was carried out by Photo Diode Array detector at 216 nm. Chromatographic resolution of the LAF was achieved within 4.6 min by using mobile phase Methanol and 5 mM Di-Potassium Hydrogen Phosphate Buffer (pH 9.5) (80:20, v/v), flow rate was 1.0 mL/min. Calibration curve was linear with correlation coefficient of 0.9996 in the range of 50-1000 ng/mL, Limit of Detection (LOD) and Limit of Quantitation (LOQ) were 10 ng/mL and 30 ng/mL respectively, intra and inter-day deviations were lower than 3.92% and 3.98% respectively. The overall mean recovery of LAF was 94.57%. No any endogenous constituents were found to interfere at retention time of the analyte. This new RP-HPLC method was successfully validated and may be applied to conduct bioavailability & bioequivalence studies of LAF.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.