Abstract

Chloramphenicol (CAP) is a toxic substance for human health, and detection of CAP residues in milk is necessary. However, most of the traditional CAP detection methods including high performance liquid chromatography-tandem mass spectrometry (HPLC-MS) and enzyme-linked immunosorbent assay (ELISA) are time-consuming and complicated. Herein, an automated microfluidics system for CAP detection in milk was developed. The residual CAP of multiple milk samples was quantitatively detected via competitive immunoassay in a single microfluidic chip simultaneously and automatically, and the reliability of the method was confirmed by flow cytometry. Completion of the detection by the system required less than 20 min and the cost for the detection of ten samples was about US$2.5. The limit of detection was 0.05 µg L−1, and the recovery rate of CAP in milk ranged from 91.3% to 105.5%. The microfluidic system developed in this study exhibited considerable potential in the point-of-care testing (POCT) of CAP in milk.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.