Abstract

Cyclic redistribution of air within the cuff during respiratory pressure changes creates a self-sealing mechanism which allows tracheal sealing, despite tracheal airway pressure being above baseline cuff inflation pressure. The aim of the present study was to investigate the effect of continuous automated cuff pressure regulation on tracheal sealing during cyclic respiratory pressure changes. In vitro tracheal sealing was studied in four different high volume-low pressure (HVLP) tracheal tube cuffs size internal diameter 8.0 and 5.0 mm in combination with a conventional pressure manometer and two different automated pressure controllers (VBM Cuff Controller; Cuff Pressure Control Tracoe). Experiments were performed at 10, 15, 20, and 25 cm H(2)O cuff pressure during intermittent positive pressure ventilation with peak inspiratory pressures of 20 and 25 cm H(2)O. Air leakage was assessed spirometrically. Experiments were performed four times with each tube brand and size with two exemplars of each of the three cuff pressure controllers. Owing to immediate cuff pressure correction, tracheal sealing at cuff pressure below inspiratory pressure was reduced in most of the tracheal tube cuffs, except in those with reduced sealing characteristics when using the Pressure Control Tracoe compared with the conventional pressure manometer and the VBM Cuff Controller. Tracheal sealing with the Pressure Control Tracoe comparable with the other two devices was only achieved at cuff pressures of 20 and 25 cm H(2)O. Automated cuff pressure controllers with rapid pressure correction interfere with the self-sealing mechanism of high sealing HVLP tube cuffs and reduce their improved sealing characteristics.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call