Abstract

Anti-drug antibodies can interfere with the activity of anti-tumor necrosis factor (TNF) agents by increasing drug clearance via direct neutralization. The presence of anti-drug antibodies is clinically relevant when trough drug concentrations are undetectable or sub-therapeutic. However, traditional immunoassay is not easily and rapidly accessible, making the translation of the results into treatment adjustment difficult. The availability of a point-of-care (POC) test for therapeutic drug monitoring (TDM) might represent an important step forward for improving the management of inflammatory bowel disease (IBD) patients in clinical practice. In this pilot study, we compared the results obtained with POC tests with those obtained by enzyme-linked immunosorbent assay (ELISA) in a group of IBD patients treated with Infliximab (IFX). We showed that POC test can reliably detect presence of antibody-to-IFX with 100% of specificity and 76% sensitivity, in strong agreement with the ELISA test (k-coefficient = 0.84).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.