Abstract

Moisture analysis by gas chromatography was performed on tuna flesh, beef, and chicken meat. Weighed quantities of meat were homogenized in anhydrous methanol, ethanol, or isopropanol with an ultrasonic disintegrator. A complete extraction was accomplished by these solvents. The extracts were analyzed by gas chromatography using a Porapack Q column and thermal conductivity detector. The water content was quantitated by calculating the peak areas of water/solvent (A1/A2), and comparing it with a standard plot of A1/A2 for known weight of water. The moisture values obtained by this technique were in good agreement with those of the official AOAC air oven drying procedure. The time required for such analysis was about 5minutes compared to 16-18 hours in AOAC method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.