Abstract

Abstract A fast mass analysis method using digital ion trap technology was reported. The mass analysis was realized by scanning the frequency of digital RF voltage and dipole ion resonance voltage during ion injection period without the conventional ion cooling, ion resonance ejection or ion cleaning process. It can decrease about three fourth of traditional analysis time using ion trap mass spectrometry. The ion signal was optimized through the modification of frequency scanning rate, ion gate voltage and other experimental parameters. For example, an optimal mass spectroscopy of reserpine ( m / z 609) and arginine ( m / z 174) were obtained when the ion gate voltage was 9 V and the digital RF frequency was scanned from 1 MHz to 400 kHz at rate of 2385 Th s −1 . The results indicated that the obtained mass spectra were in very good agreement with conventional analytical method using ion trap mass analyzer.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call