Abstract

AbstractPolyacrylamide gel isoelectric focusing was carried out using 250 μm thick gels backed with polyester films at voltage gradients up to 460 V/cm. Peltier cooling was employed with berrylium oxide heat transfer plates to achieve adequate cooling. Visualization of low protein concentrations was obtained by the use of a highly sensitive silver staining technique. Results indicated that over separating distances of 5.4 cm 150 to 200 bands should be resolved and that considerable differences between two commercial carrier ampholytes exist. Pharmalyte carrier ampholytes from Pharmacia have better continuity and conductance over the range of pH 3–10 but require much higher field strengths to achieve maximal separation than do LKB Ampholines over a similar pH range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.